Quality Assurance of the Segmented Mono-Isocentric Technique in Breast Radiotherapy

Loading...
Thumbnail Image

Date

Journal Title

Journal ISSN

Volume Title

Publisher

Setif 1 Unuversity Ferhat Abbas . Faculty of Sciences

Abstract

This study evaluated diode-based in-vivo dosimetry and EPID-based portal dosimetry for quality assurance in breast cancer radiotherapy using the 3D conformal Field-in-Field (FIF) technique. Diode calibration and correction factors were successfully established according to the IAEA TRS-398 protocol. In-vivo dosimetry measurements showed good agreement between measured and calculated entrance doses, with most deviations remaining within the accepted ±10% tolerance for heterogeneous breast irradiation. Phantom measurements demonstrated improved agreement under homogeneous conditions. EPID portal dosimetry showed excellent agreement between predicted and measured dose distributions, with high gamma passing rates using the 3%/3 mm criterion. The combined use of diode and EPID dosimetry proved to be an effective and reliable quality assurance strategy for improving treatment verification and patient safety in breast radiotherapy.

Description

تناولت هذه الدراسة تقييم القياس الجرعي داخل الجسم باستخدام الديود، إضافةً إلى القياس الجرعي البوابي المعتمد على جهاز EPID ، كوسائل لضمان الجودة في العلاج الإشعاعي لسرطان الثدي باستعمال تقنية العلاج الإشعاعي ثلاثي الأبعاد بتقنية ال Field-in-Field (FIF). تم إجراء معايرة الديود وتحديد عوامل التصحيح وفق بروتوكول الوكالة الدولية